Integra
Miltex
®
Dental Aspirating Syringes
Directions for Use
EC REP
Manufacturer
Integra York PA, Inc.
589 Davies Drive, York, PA 17402 USA
866-854-8300 USA
717-840-2763 outside USA 717-840-9347 fax
integralife.com/integra-miltex
Description
Integra Miltex® Aspirating Syringes are used in conjunction with
a needle and a 1.8cc pre-filled anesthetic cartridge (not supplied by
Integra™ Miltex®). Aspirating syringes are available in Type CW
(Cook-Waite) and Type A (Astra). The syringes are reusable and are
provided clean and supplied non-sterile.
Indications
Integra Miltex® Aspirating Syringes are indicated to be used in
conjunction with anesthetic needles and anesthetic cartridges for
injection of anesthetic solutions in the oral cavity.
Contraindications
None known.
Warnings
To avoid needle sticks, re-cap needle after injection
has been given.
To avoid leakage during an injection, ensure that needle is
straight as it penetrates the cartridge diaphragm.
Precautions
Syringes must be sterilized prior to use. See sterilization
instructions.
Hubs cannot be removed. Do not attempt to remove hub
from syringe for sterilization or for any reason as damage to
the syringe will occur.
Do not mix metals when sterilizing syringes. Sterilizing
syringes with other non chrome-plated brass
instruments may cause damage to syringes.
Do not use cold sterilants, as they will react to the chrome
plating of the syringe.
Keep needle capped while affixing or removing from syringe.
Instructions
Loading Instructions
1. Retract piston rod fully.
2. Insert 1.8cc cartridge, plunger end first, into syringe barrel.
3. Hold cartridge firmly in place. Push piston rod forward with
a quick jab.
4. Affix needle to aspirating syringe.
5. Expel a few drops of solution to clear needle.
6. Aspirate before injecting solution after inserting needle
into tissue.
7. After positioning the needle, pull back on thumb ring to aspirate.
Unloading Instructions
1. Remove needle from syringe.
2. Withdraw piston rod to clear cartridge for removal.
3. Tilt syringe to expel cartridge.
Cleaning Instructions
• For best and long service life, syringes should be processed
according to Integra™ Miltex® Care and Cleaning guidelines.
• Remove all traces of blood, saliva, and anesthesia solution.
Remove paraffin that has accumulated on the harpoon.
Sterilization Instructions
Syringes are supplied non-sterile and must be sterilized
prior to each use.
Syringes should not be sterilized with other instruments to
avoid syringe damage.
Before each use, sterilize and inspect instrument. Sterilize
syringes according to an approved National/International
Standard. Integra™ Miltex® syringes have been tested with
no adverse effects under the following sterilization parameters
in a standard, gravity steam, sterilizer:
- Temperature: 250ºF (121ºC)
- Exposure Time: 30 minutes
- Dry Time: 20 minutes
Do not use cold sterilants, as they will react to the chrome
plating of the syringe.
Symbols
Symbols may be used on some package labeling for easy identification.
Integra and the Integra logo are trademarks of, and Miltex is a registered
trademark of, Integra LifeSciences Corporation or its subsidiaries.
©2011 Integra LifeSciences Corporation. All rights reserved.
DENTASPDFU Rev. B 06/13
EC REP
Manufacturer
Authorized Representative
in the European Community
Catalog number
Lot number
Caution: see warnings or precautions
Product complies with requirements of directive
93/42/EEC for medical devices
Federal (USA) law restricts this device to sale by or
on the order of a physician or practitioner
Consult instructions for use and website address
ENGLISH
Integra LifeSciences Services (France) SAS Immeuble Sequoia 2
97 allée Alexandre Borodine
Parc Technologique de la Porte des Alpes
69800 Saint Priest–France
Telephone: 33 (0) 4 37 47 59 10
Fax: 33 (0) 4 37 47 59 29
Seitenansicht 0
1 2 3 4 5 6

Inhaltsverzeichnis

Seite 1 - Dental Aspirating Syringes

Integra™ Miltex®Dental Aspirating Syringes Directions for UseEC REPManufacturerIntegra York PA, Inc.  589 Davies Drive, York, PA 17402 USA866-854-83

Seite 2 - Jeringas aspiradoras dentales

EC REPFabricanteIntegra York PA, Inc.  589 Davies Drive, York, PA 17402 EE.UU.866-854-8300 EE.UU.  717-840-2763 fuera de EE.UU.  717-840-9347

Seite 3 - Mode d’emploi

ESPAÑOLEC REPFabricantIntegra York PA, Inc.  589 Davies Drive, York, PA 17402 États-Unis866-854-8300 USA  717-840-2763 en-dehors des États-Unis

Seite 4 - Dental-Ansaugspritzen

GERMANEC REPHerstellerIntegra York PA, Inc.  589 Davies Drive, York, PA 17402 USA866-854-8300 USA  717-840-2763 außerhalb der USA  717-840-934

Seite 5 - Istruzioni per l'uso

GERMANITALIANOEC REPProduttoreIntegra York PA, Inc.  589 Davies Drive, York, PA 17402 USA866-854-8300 USA  717-840-2763 negli altri paesi  717

Seite 6 - Gebruiksaanwijzing

DEUTSCHEC REPFabrikantIntegra York PA, Inc.  589 Davies Drive, York, PA 17402 VS866-854-8300 VS  717-840-2763 buiten VS  717-840-9347 faxinteg

Kommentare zu diesen Handbüchern

Keine Kommentare